For a CO2 accuracy specification written ±(A ppm + B% of reading), multiply the stated reading by B/100, add A, and retain the result as a plus-or-minus allowance in ppm. For the hypothetical specification ±(50 ppm + 3% of reading), a reading of 1,000 ppm gives 50 + 30 = 80 ppm. The expression is ±80 ppm under the conditions attached to that specification.
The arithmetic does not establish a true concentration, a confidence interval, or a calibration result. It translates a manufacturer's claim into a useful unit. To decide whether a monitor meets an application requirement, also check the concentration range, environmental conditions, calibration assumptions, output path, and verification method. The same monitor can have a broad measurement range and a narrower range over which its accuracy is specified.

Conceptual instrument comparison scene. All numerical examples below are constructed examples, not measurements of the pictured equipment.
Read the entire expression before calculating
The fixed ppm term remains the same throughout a stated range. The reading term increases with concentration. Neither is optional when the formula uses a plus sign. A useful calculation worksheet has separate columns for concentration, fixed contribution, proportional contribution, total allowance, and applicable conditions. This makes a misunderstood percent sign easier to detect than a single final result.
Use the concentration named by the source. Some verification procedures evaluate the allowance at the reference concentration, while a product specification may explicitly say percent of measured value. These conventions are close at small errors but are not identical. Record the convention and follow the manufacturer's verification method rather than silently switching between them. In the examples here, C is the displayed or measured reading unless explicitly stated otherwise.
At 400 ppm, the hypothetical 3% term contributes 12 ppm, giving 62 ppm overall. At 2,000 ppm, it contributes 60 ppm, giving 110 ppm overall. At 5,000 ppm, it contributes 150 ppm, giving 200 ppm overall. The final allowance increases, even though the same formula applies throughout. The relative allowance decreases because the fixed contribution becomes a smaller fraction of the concentration.
| Reading C | Fixed contribution | 3% contribution | Total A + 0.03C | Total divided by reading |
|---|---|---|---|---|
| 400ppm | 50ppm | 12 ppm | 62 ppm | 15.5% |
| 1,000ppm | 50ppm | 30ppm | 80ppm | 8.0% |
| 2,000ppm | 50ppm | 60ppm | 110 ppm | 5.5% |
| 5,000ppm | 50ppm | 150 ppm | 200ppm | 4.0% |
The last column is a calculated relative allowance for this example. It is not a new independent specification. Near zero, dividing by the reading becomes unhelpful or undefined, while the fixed term still has meaning. Do not force a percentage comparison at zero concentration.
The two contributions are equal when A = bC, where b is the percentage expressed as a fraction. Here C = 50/0.03 = 1,666.7 ppm. Below that value, the fixed term contributes more; above it, the reading term contributes more. This crossover is useful when selecting comparison points, but it is not a physical transition in the sensor.

The two contributions are added in the stated specification. The blocks are conceptual and do not encode measured values.
Plus, or, and full scale give different answers
A phrase such as “50 ppm or 3% of reading, whichever is greater” is a different rule. At 1,000 ppm, the two candidates are 50 ppm and 30 ppm, so the larger is 50 ppm. The additive expression gives 80 ppm. At 2,000 ppm, the larger candidate is 60 ppm, while the additive expression gives 110 ppm. Copying the same numerical terms without their connecting words changes the claim.
Percentage of full scale is also different. A hypothetical ±3% full-scale allowance on a 5,000 ppm range is ±150 ppm throughout that range. At 1,000 ppm, ±3% of reading is only ±30 ppm. Changing a transmitter's configured output span may or may not change a manufacturer's specified full-scale basis; the relevant documentation must identify which span is meant.
Percentage points require another distinction. A change from 1.00% CO2 to 1.03% CO2 is 0.03 percentage points, equal to 300 ppm. It is a 3% relative increase from 1.00%. Meanwhile, 3% of a 1,000 ppm reading is 30 ppm. The percent symbol alone does not identify the arithmetic. Write down the base quantity before doing a conversion.
Translate a product claim without expanding its scope
の Sensirion SCD30 データシート, version 1.0, May 2020, specifies ±(30 ppm + 3% of measured value) over 400–10,000 ppm. Its table and footnotes also identify reference conditions and calibration qualifications. At a measured value of 1,000 ppm, the expression becomes ±60 ppm; at 5,000 ppm, it becomes ±180 ppm. Those calculations apply the published formula, not an independent evaluation of any finished product containing that module.
The same document distinguishes a wider digital measurement range from the accuracy range. A valid numerical output beyond an accuracy range does not extend that range. A complete instrument may additionally limit its display, transmitter scaling, approved application, or environmental conditions. For procurement, put the relevant range next to each calculated number so that an apparently impressive comparison cannot hide an out-of-range point.
の Murata accuracy FAQ gives separate actual-performance and specification expressions. This is a useful reminder that a typical or observed performance statement cannot automatically replace a formal specification. Identify the exact model, current document revision, product status, and conditions before treating a general FAQ statement as a contractual acceptance limit.
Product pages can summarize several channels in a compact card. For example, the DwyerOmega Series-CO2 page includes CO2, humidity, and temperature information. A short resolution field without a channel identifier should not be assumed to describe CO2. The detailed CO2 accuracy line is the relevant place to start, followed by the model's full manual and data table.
A comparison can reverse with concentration
Consider two hypothetical instruments, both otherwise suitable for the same conditions. Instrument A is specified at ±(30 ppm + 3% of reading). Instrument B is specified at ±(50 ppm + 2% of reading). At 500 ppm, A gives 45 ppm and B gives 60 ppm. At 3,000 ppm, A gives 120 ppm and B gives 110 ppm. The better numerical allowance changes with concentration.
Set 30 + 0.03C = 50 + 0.02C to find the crossover. Subtracting terms gives 0.01C = 20, so C = 2,000 ppm. Both expressions give 90 ppm there. This is a comparison of two assumed formulas, not a ranking of brands, calibration systems, or overall reliability. A narrower allowance is useful only if its conditions and scope fit the intended measurement.
| Hypothetical reading | A: 30 ppm + 3% | B: 50 ppm + 2% | Smaller formula allowance |
|---|---|---|---|
| 500ppm | 45 ppm | 60ppm | あ |
| 1,000ppm | 60ppm | 70ppm | あ |
| 2,000ppm | 90 ppm | 90 ppm | Equal |
| 3,000 ppm | 120 ppm | 110 ppm | B |
Compare at concentrations that matter to the application rather than selecting only the crossover or an easy ambient point. A room-monitoring application and a higher-concentration process application can reasonably emphasize different points. Avoid treating a ventilation product's ppm expression as evidence that it is suitable for an unrelated process gas or safety function.
Keep allowance, measured deviation and uncertainty separate
Suppose a monitor indicates 1,080 ppm while a suitable reference reports 1,000 ppm under stable, matched conditions. The observed signed deviation is +80 ppm. This subtraction describes the comparison. It does not by itself prove that the monitor's error relative to the true measurand is exactly +80 ppm, because the reference and the comparison arrangement also contribute uncertainty.
For the hypothetical ±(50 ppm + 3% of reading) rule, evaluating at the monitor's 1,080 ppm gives 82.4 ppm. Evaluating at the 1,000 ppm reference gives 80 ppm. This is a deliberately marginal example: the convention matters near a decision boundary. A test method should specify the evaluation point and decision rule before the result is known. Choosing whichever convention passes a result is not a defensible verification method.
の 計量学の国際語彙 provides the terminology behind error, uncertainty and precision. An engineering worksheet should keep the observed comparison, published allowance and comparison uncertainty in separate fields. Doing so prevents a tolerable disagreement from becoming a false claim of exact accuracy.
A manufacturer's additive formula is not an instruction to calculate the root-sum-square of its two terms. For 50 ppm and 30 ppm, the specified sum is 80 ppm; the root-sum-square would be about 58.3 ppm and would change the claim. Statistical combination belongs to an explicitly defined uncertainty model with justified distributions and dependencies, not to casual rewriting of an accuracy line.
Likewise, do not add the reference uncertainty to the product allowance and declare the larger result a universal pass band. Some decision rules use guard bands or other treatment of uncertainty. The applicable contractual or laboratory procedure determines the rule. A useful local record can state “observed deviation +80 ppm; allowable magnitude 80 ppm at the reference point; uncertainty and decision rule recorded separately” without claiming more than the evidence supports.

Co-location only supports a comparison when gas exposure, timing, environmental conditions and reference suitability are controlled.
Two monitors can disagree while each fits its allowance
At an assumed common concentration of 1,000 ppm, two instruments with ±80 ppm allowances could hypothetically indicate 920 and 1,080 ppm. The disagreement would be 160 ppm. This is a simple bound illustration assuming the same allowance basis and conditions. It is not the predicted distribution of differences between real instruments.
Agreement also has limits as evidence. Two monitors that read 1,050 ppm could share a calibration assumption, environmental compensation problem, or exposure bias. Matching displays therefore do not establish traceability. A third consumer monitor can help locate a gross inconsistency, but it becomes a reference only when its suitability and uncertainty for the comparison are established.
Check the conditions that can invalidate a neat calculation
Temperature, pressure, humidity, calibration status, warm-up, sampling and housing design can all determine whether a stated specification applies. Their effects should be handled according to the model documentation. Do not automatically add every published sensitivity coefficient to an accuracy term: some effects may already be included within a stated condition range, while others are explicitly additional.
の SCD30 デザインイン ガイド emphasizes gas exchange, separation from external heat sources and protection against problematic airflow. These are integration conditions, not decorative installation advice. A module installed beside a warm processor or behind an unsuitable enclosure opening is a different measurement arrangement from a bare module under its reference conditions.
Pressure deserves particular attention when a monitor is moved between sites or connected to a sampling system. A gas sensor may require an external absolute-pressure value, use a configured assumption, or include an onboard measurement. “Pressure compensated” does not establish which of these applies. Check the value actually used and whether it represents the gas at the measurement cell.
Humidity basis can create another comparison problem. A reference that reports dry-gas mole fraction and a monitor exposed to humid air may not report the same measurand. Correcting the difference requires measured conditions and an appropriate method, not a generic offset. Similarly, a cylinder's nominal concentration alone is insufficient without its certificate, balance gas, validity and delivery suitability.
| Condition to record | なぜ解釈が変わるのか | What to obtain |
|---|---|---|
| Accuracy concentration range | Output range can be wider | Exact model table and footnotes |
| Temperature and pressure | Compensation and residual effects are conditional | Actual values and active settings |
| Calibration state | Adjustment can conceal an as-found error | Last action, mode and before/after readings |
| Sampling arrangement | Different gas or delay can mimic disagreement | Positions, tubing and stabilized interval |
| Output used | Display and analog path can differ | Digital reading, scaling and receiving channel |
| Reference basis | Dry/wet or balance-gas mismatch changes comparison | Certificate and measurement definition |

Environmental inputs belong to the measurement model. Their presence on a bench does not demonstrate that a finished monitor uses them correctly.
Build an acceptance worksheet around the application
Start with the quantity being measured and the decision it will support. Write the relevant concentration interval and environmental envelope. Then identify the exact product specification that applies to that interval. A single spreadsheet cell labelled “accuracy” cannot preserve all of this information; a small structured table is more useful.
For each point, retain the original formula as text, calculate the fixed and proportional terms separately, and record whether the percentage is based on reading, reference or full scale. Include units in headings and in any exported report. Keep the unrounded calculation internally, while reporting the final allowance with sensible precision consistent with the specification and procedure.
Consider a hypothetical monitoring requirement that permits a magnitude of deviation no greater than 100 ppm at selected check points. A candidate specified at ±(50 ppm + 3% of reading) gives 80 ppm at 1,000 ppm and 110 ppm at 2,000 ppm. On the formula alone, it fits the first numerical requirement and exceeds the second. This tells the buyer where further evidence or a different product may be needed; it does not establish actual failure of an untested device.
Do not replace this review with an average allowance across the range. An average could hide the point where the application is most demanding. Also distinguish a requirement for absolute concentration from one for small concentration differences. Subtracting two readings can introduce contributions from both measurement chains, so the single-instrument allowance is not automatically the allowance for the difference.
Preserve the observation before adjustment
Record the as-found result before changing calibration, compensation, filters or scaling. A post-adjustment result answers whether the adjusted system meets the chosen check; it cannot reconstruct the error present before the work. Include instrument identity, firmware, settings, reference identity, environmental values, stabilization criterion, raw readings and calculation convention.
Repeat a check when the original arrangement did not provide stable or representative exposure. Document why the first result was invalid rather than deleting it. If a repeat differs, identify what changed: gas delivery, equilibration, pressure input or instrument configuration. Otherwise, the worksheet becomes a collection of selected passing numbers rather than evidence about the instrument.
For recurring verification, use the same comparison points where practical. Changing all points, the reference, the software filter and the exposure arrangement together makes a later difference difficult to interpret. A controlled record helps separate a real instrument change from a changed method. The procedure should still be revised when the original setup is unsuitable; reproducibility of a flawed method is not the objective.

A useful acceptance record retains the original specification, actual conditions, measured deviation and decision rule together.
Common calculation traps and how to catch them
Entering 3 instead of 0.03 multiplies a reading by three rather than by three percent. A simple independent check is to calculate 1% first: at 1,000 ppm, 1% is 10 ppm, so 3% must be 30 ppm. This mental check catches both decimal errors and spreadsheet cells that already apply percentage formatting.
Mixing ppm and percent concentration is another trap. One percent concentration corresponds to 10,000 ppm. If a spreadsheet stores a 1,000 ppm reading as 0.1% concentration, its numerical representation must be handled consistently before combining with a term expressed in ppm. Convert both contributions into the same unit before addition.
Using the configured analog span as the denominator without checking the claim can produce a much larger error. A transmitter may have a physical measurement range, an accuracy range and a user-configured output span. They answer different questions. Read the source's definition of full scale instead of inferring it from the screen or control-system configuration.
Rounding early can affect a borderline result. If a proportional contribution is 32.4 ppm, preserve it until the final reported allowance is calculated. The acceptance procedure should describe rounding and boundary handling. A rounded display can conceal a small underlying difference, so retain raw digital or logged values when the procedure permits and the interface is understood.
Finally, a published formula is not a calibration correction. Subtracting 80 ppm from every reading because the allowance is ±80 ppm has no technical basis. The formula describes a permitted or stated magnitude under conditions; it gives neither the sign nor the actual magnitude of an individual instrument's deviation. Apply a correction only from a justified calibration model.
Review the complete instrument, not only its sensing element
A product built around a documented sensing module has additional design choices. The enclosure controls gas access, the power supply affects operating conditions, the software applies settings, and the output path presents the result. A module's data table is valuable evidence about the component. It does not automatically describe every finished monitor that uses it.
For an OEM comparison, ask which conditions were verified after integration. Was the final enclosure fitted? Were its usual processor, display and communications functions operating? Which compensation values were applied? Was the output taken from the module, the front panel or a building controller? These questions connect the published formula to the measurement that the application will actually use.
Consider a hypothetical integrated monitor whose sensing module meets an 80 ppm allowance at a check point. If the receiving system has an independently identified offset, the application may observe a larger deviation. The correct response is to characterize the full path. Assigning the whole discrepancy to the gas module or subtracting a guessed fixed allowance conceals the actual issue.
An acceptance specification can state both a component requirement and a complete-system requirement when necessary. Keep their scopes distinct. A supplier may reasonably provide a module report while an integrator supplies the final monitor evidence. The reader needs to know which measurement arrangement each report covers and whether the intended installation remains within that arrangement.
Keep calculations reproducible when a specification changes
Record the document revision used in a comparison. If a later revision changes the expression, range or conditions, preserve the earlier calculation rather than silently overwriting it. A procurement decision made under one claim and a later maintenance check under another need an explicit relationship. The change may reflect an expanded product range, a revised test method or a corrected summary; investigate instead of assuming performance changed physically.
A reproducible worksheet stores inputs separately from formulas. Put A, b, the relevant reading and any full-scale basis in named columns. Include an independent arithmetic check at one or two simple concentrations. A copied spreadsheet cell can retain the wrong span or percentage base even when its displayed result appears plausible.
For example, using 2 instead of 3 as the percentage coefficient at 2,000 ppm gives 90 ppm rather than 110 ppm for the hypothetical 50 ppm fixed-term specification. Both results look reasonable in isolation. Keeping the original text beside the input fields makes the discrepancy detectable during review. The record should show where each coefficient came from rather than relying on memory.
Avoid formatting a computed allowance as an observed error. Label it “calculated specification allowance” and give the evaluation concentration. Label a test subtraction as “observed reference deviation.” Label uncertainty with its stated coverage information when available. These headings make the report understandable to a later reviewer who did not watch the comparison being performed.
よくある質問
Does ±80 ppm mean the real concentration is always within 80 ppm of the display?
It describes the translated specification under its stated scope and conditions. It is not automatically an absolute guarantee for every device, environment or time, and it is not a probability statement. Consult qualifications concerning production population, calibration and operating conditions. A measurement result with uncertainty needs its own documented comparison or calibration basis.
Can I ignore the fixed term at high concentrations?
Only if the applicable documentation explicitly permits a different expression there. The fixed term may become a smaller proportion of the total, but it remains part of an additive formula. For the worked 5,000 ppm example it is still 50 of the 200 ppm total, so dropping it would understate the stated allowance by one quarter.
Does one-ppm display resolution imply one-ppm accuracy?
No. Resolution is the output increment; accuracy concerns agreement with the measurand under specified conditions. A monitor can show small increments while having a much larger systematic deviation or combined allowance. More digits can be useful for logging trends, but they do not replace reference verification.
Should I choose the smallest ppm number in a product comparison?
Compare the entire formula at relevant points, then compare the conditions and the complete instrument. A smaller fixed term can accompany a larger proportional term. Also check whether one claim is typical, one is a maximum, one applies to a module and another to a finished monitor. The numbers become comparable only after those differences are resolved.
Is outdoor air a sufficiently known reference for this calculation?
Outdoor air provides an environmental observation, not an automatically certified concentration. Local sources, mixing, sampling location and time affect it. It may be part of a manufacturer's specified baseline procedure, but that procedure is different from demonstrating absolute accuracy with a suitable reference. Do not silently substitute an assumed outdoor number for a documented comparison value.
What should I ask a supplier when the wording is ambiguous?
Request the exact model specification and revision, the meaning of the percentage base, the connecting rule between terms, the accuracy range, the environmental conditions and the recommended verification method. Ask whether the claim covers the complete monitor and output you will use. A written clarification is more useful than guessing from a shortened catalogue line.
出典と詳細情報
- Sensirion SCD30 datasheet, version 1.0, May 2020: model-specific expression, range and qualifications.
- Sensirion SCD30 design-in guidelines, February 2019: enclosure, heat and airflow conditions.
- Murata CO2 sensor accuracy FAQ: distinction between stated performance expressions.
- DwyerOmega Series-CO2: channel-specific product specification example.
- JCGM 200:2012、計量学の国際語彙: terminology for interpreting measurement claims.
- Atmospheric pressure and its measurement: background for understanding a pressure condition in a gas measurement.
コメント (11)
[…] Twitter LinkedIn Google + 電子メール 2026-10-12 コメントはありません センサー 1 […]
[…] How to read a CO2 monitor accuracy specification: interpreting a combined ppm and percentage allowance. […]
[…] Reading a CO2 monitor accuracy specification: keeping a stated allowance distinct from observed comparison results. […]
[…] humidity basis, thermal state and response time can still limit the comparison. The guide to reading a CO2 accuracy specification explains why the conditions attached to a specification […]
[…] of full scale, percentage points, maximum error and typical behavior are different. The guide to reading accuracy specifications uses CO2 examples, but its distinction between error terms and conditions is also useful […]
[…] a device specification, see how fixed and percentage CO₂ accuracy terms apply at a given concentration. That arithmetic helps identify whether a transmitter's documented performance fits the […]
[…] To evaluate each device on the correct basis before taking a difference, see how to interpret a CO₂ monitor accuracy specification. […]
[…] CO₂ monitor accuracy guide explains how to compare stated tolerance terms at the concentrations that will affect a decision. […]
[…] differently, the observed difference may partly reflect their response rather than the room. The CO₂ monitor accuracy guide explains how to evaluate the separate specification terms involved in interpreting two […]
[…] the device or configuration is replaced, preserve the prior trend period and verify the new state. Sensor1stop's guide to CO₂ monitor accuracy specifications explains how to read a manufacturer accuracy statement when comparing a field monitor or […]
[…] or temperature effects. Read the manufacturer’s definitions before adding or comparing terms. The CO₂ monitor accuracy guide works through fixed and concentration-dependent terms for that specific instrument category. A […]